DL-301-IP65/DL-302-IP65/DL-303-IP65/
DL-304-IP65/DL-305-IP65/DL-306-IP65/
DL-307-IP65/DL-308-IP65

CO/CO2/Temperature/Humidity/Dew Point Data Logger

Introduction
The DL-300-IP65 series of Data Logger devices can be used to record CO, CO2, HCHO, TVOC, NH3, H2S, Temperature, Humidity and Dew Point information, including date and time stamps, and are able to store up to 450,000 downloadable records. Real-time data can be accessed from the DL-300-IP65 series Data Logger from anywhere and at any time using the free Windows software, the iOS App or the Android App, as long as they are connected to the same local network as the Data Logger. Support is provided for popular industrial protocols such as DCON, Modbus RTU, and Modbus TCP, as well as the emerging machine-to-machine (M2M)/IoT (Internet of Things) connectivity protocol-MQTT. The DL-300-IP65 series Data Logger can be connected via widely used communication interfaces including RS-485, Ethernet and PoE, meaning that the device can be easily integrated into existing HMI or SCADA systems, and is easy to be maintained in a distributed control system. The IP65 version of DL-300-IP65 series is designed for industrial applications in harsh environments that provides IP65 grade protection approval. The rugged RJ-45 ensures tight, robust connections, and guarantees reliable operation, even for applications that are subject to high vibration and shock.

Request For Quotation


    Description

    Features
    • Able to record CO, CO2, HCHO, TVOC, NH3, H2S, Temperature, Humidity, and Dew Point Measurements
    • Non-dispersive Infrared (NDIR) CO2 Sensor
    • Electrochemical Sensor
    • 2.8″ LCD Touch Screen
    • Up to 450,000 records with date and time stamps
    • Touch-screen and Web-based Configuration Interface
    • Simple and Powerful Software Utility, iOS APP and Android App Included
    • Able to Display Multilingual Messages on the Screen
    • Supports the DCON, Modbus RTU/TCP, and MQTT Protocols
    • Includes RS-485/Ethernet/PoE Communication Interfaces
    • Relay Output for Audible/Visual Alarm or IAQ Device Control
    • Includes redundant power inputs: PoE (IEEE 802.3af, Class 1) and DC input
    • IP 65 Protection Approval
    Applications
    Mushroom Farm
    Mushroom growth yield requires long time to monitor the environment of CO2 concentration, temperature and humidity values, and its growth environment is high humidity condition. The general equipment cannot work in such environment, but the DL-302-IP65 can be used to record CO2, temperature, humidity and dew point information with date and time stamps in the operating temperature from -20˚C to +50˚C, and are able to store up to 450,000 downloadable records. The WISE-5231-3GWA features timer and schedule functions. It allows user to schedule specific date or time for control logic execution, or perform specific tasks such as time delay. With calendar user interface provided, schedule setting can be more efficient and flexible. This does not require programming, only some simple settings will be able to achieve the farm production control and quality control, production resume … and so on.

    Pig Farm
    Ammonia and Hydrogen sulfide are two of the most harmful gases in animal farming environment. They will directly affect animals’ growth and immunity. Pig farm requires long-term immediate monitoring the environment of NH3 concentration, temperature and humidity values. The general equipment cannot work in such environment, but the DL-305-IP65 can be used to record NH3, temperature, humidity and dew point information with date and time stamps in the operating temperature from -20°C to +50°C, and are able to store up to 450,000 downloadable records. The WISE-5231-3GWA features timer and schedule functions. It allows user to schedule specific date or time for control logic execution, or perform specific tasks such as time delay. With calendar user interface provided, schedule setting can be more efficient and flexible. This does not require programming, only some simple settings will be able to achieve the farm production control and quality control, production resume … and so on.

    Specifications

    Models DL-301-IP65 DL-302-IP65 DL-303-IP65 DL-304-IP65 DL-305-IP65 DL-306-IP65 DL-307-IP65 DL-308-IP65
    System
    CO Alarm Yes Yes
    CO2 Alarm Yes Yes
    HCHO Alarm Yes Yes
    TVOC Alarm Yes Yes
    NH3 Alarm Yes
    H2S Alarm Yes
    RH/Temperature Alarm Yes
    Real Time Clock Yes
    Data Logger Yes, up to 450,000 records
    Relay Output PhotoMos Relay × 4, 100 VDC @ 1 A
    Communication
    RS-485 Port Baud Rate = 1200 ~ 115200 bps
    Ethernet Port 10/100 Base-TX, 8-Pin RJ-45 x1(Auto-negotiating, Auto-MDI/MDIX, LED indicators)
    Security IP filter (whitelist) and Password (web)
    Protocol Modbus/RTU(RS-485), Modbus TCP(Ethernet) and MQTT(Ethernet)
    Dual Watchdog Yes, Module (2.3 sec), Communication (Programmable)
    LED Indicators
    PWR Green for normal operation
    Link Green for the Ethernet-linked
    S1 ~ S4 Red for an alarm condition
    Main Machine Interface
    LCD 2.8 TFT (Resolution 240 x 320 x 16), Defective Pixels <= 3
    Backlight Life 20,000 hours
    Brightness 160 cd/m2
    Touch Panel Yes
    Electrical
    Powered from Terminal Block +12 to +48 VDC
    Powered from PoE IEEE 802.3af, Class 1 (48 V)
    Power Consumption PoE 2.9 W (Max.) 3.0 W (Max.) 3.1 W (Max.) 2.0 W (Max.) 2.9 W (Max.) 2.9 W (Max.) 2.1 W (Max.) 2.0 W (Max.)
    Non-PoE 2.7 W (Max.) 2.8 W (Max.) 2.8 W (Max.) 2.0 W (Max.) 3.0 W (Max.) 3.0 W (Max.) 2.1 W (Max.) 2.0 W (Max.)
    Mechanical
    Installation DIN-Rail or Wall Mounting
    Dimensions (L x W x H) 170 mm x 117 mm x 60 mm
    Environment
    Operating Temperature -20 to +50°C
    Storage Temperature -30 to +75°C
    Humidity 10 to 90% RH, Non-condensing
    Protection Class IP65

     

    I/O Specifications

    Models DL-301-IP65 DL-302-IP65 DL-303-IP65 DL-304-IP65 DL-305-IP65 DL-306-IP65 DL-307-IP65 DL-308-IP65
    CO Measurement
    Range 0 to 1000 ppm
    (Electrochemical)
    0 to 1000 ppm
    (Electrochemical)
    Resolution 1 ppm 1 ppm
    Accuracy ±5% of
    measured value
    ±5% of
    measured value
    Response Time 30 sec 30 sec
    Warm-up Time 300 sec 300 sec
    CO2 Measurement
    Range 0 to 9999 ppm (NDIR)
    Resolution 1 ppm
    Accuracy ±30 ppm ±3% of measured value
    Response Time 20 sec
    Warm-up Time 60 sec
    HCHO Measurement
    Range 0 ppb to 2000 ppb
    (Electrochemical)
    0 ppb to 2000 ppb (Electrochemical)
    Resolution 1 ppb 1 ppb
    Accuracy 0 ~ 300ppb :
    ±30ppb >
    300ppb : ±10%
    0 ~ 300ppb :
    ±30ppb >
    300ppb : ±10%
    Response Time ≤60 sec ≤60 sec
    Warm-up Time 180 sec 180 sec
    TVOC Measurement
    Range 0 ppb to 60000
    ppb (MEMS
    Metal Oxide)
    0 ppb to 60000
    ppb (MEMS
    Metal Oxide)
    Resolution 1 ppb 1 ppb
    Accuracy ±15% ±15%
    Response Time 60 sec 60 sec
    Warm-up Time 180 sec 180 sec
    NH3 Measurement
    Range 0 to 100 ppm
    (Electrochemical)
    Resolution 1 ppm
    Accuracy ±5% of
    measured value
    Response Time < 40 Sec
    Warm-up Time 60 Sec
    H2S Measurement
    Range 0 to 100 ppm
    (Electrochemical)
    Resolution 1 ppm
    Accuracy ±5% of
    measured value
    Response Time < 30 Sec
    Warm-up Time 60 Sec
    Temperature Measurement
    Range -20 to +50°C
    Resolution 0.1°C
    Accuracy ±0.6°C
    Relative Humidity Measurement
    Range 0 to 100% RH, Non-condensing
    Resolution 0.1% RH, Non-condensing
    Accuracy ±5% RH, Non-condensing
    Dew Point
    Range Calculated using temperature and relative humidity
    Resolution 0.1°C

    Function

    Large 2.8″ LCD Touch Screen, with clear Color Chart to indicate the CO/CO2 Level

    Supports the MQTT Protocol for IoT Applications

    Simple and powerful utility
    The DL300 Utility can be used to configure the modules, monitor real-time data, group DL-300 modules so that the status of distribution groups can be viewed and managed. The utility also allows the log data to be downloaded and exported to a .CSV file that can then be imported into any industrystandard software or spread sheet for analysis.

    Multi-platform Remote Access Software
    Real-time data from the DL-300 Data Logger can be accessed from anywhere and at any time using the DL300 Utility, the iOS or Android App, or via a regular web browser, as long as they are connected to the same local network as the Data Logger.

    Display Messages in Multiple Languages

    The display-message-on-screen function supports multiple language character sets based on UTF-8 encoding. Either pre-configured messages or dynamic messages can be remotely displayed using Modbus commands, or a dynamic message can be sent via the web-based interface.

    Ordering Information

    DL-301-IP65 CR ~ NEW ~ Remote CO/Temperature/Humidity/Dew Point Data Logger with Ethernet/RS-485 and PoE (RoHS)
    DL-302-IP65 CR ~ NEW ~ Remote CO2/Temperature/Humidity/Dew Point Data Logger with Ethernet/RS-485 and PoE (RoHS)
    DL-303-IP65 CR ~ NEW ~ Remote CO/CO2/Temperature/Humidity/Dew Point Data Logger with Ethernet/RS-485 and PoE (RoHS)

    Datasheet

    dl-300-ip65-series.pdf

    dl-300-ip65_quick_start_v1.10.pdf

    DL-300-IP65_user_manual_english_v1.2.0.pdf

    dl-300-wf-ip65-series.pdf

    dl302_app_manual_en.pdf